Atomic Absorption Spectrometer Professional
1 year ago - Business, Industrial & Agriculture - Karachi - 160 views
Location:
Karachi
Price:
₨ --
SKYRAY (USA) Assembled in China Flame and Graphite Furnace Atomic Absorption Spectrometer Model AAS9000
AAS 9000 Integrated Atomic Absorption Spectrometer combines flame and graphite atomizers, incorporating both Flame and Graphite Furnace features. With Skyray state-of-the-art user friendly software users can switch atomizers quickly and automatically based on their needs. Users can also choose a cost effective AAS 9000-M option which comes without graphite furnace autosampler.
Features
- All 8 lamps have individual power supplies. During operation 7 lamps are preheated simultaneously while the other lamp is working-thus dramatically reducing lamp switching and preheating.
- C2H2 analysis and flame measurement: full reflection and corrected aberration
- Fully automatic operation
- 8 element lamps
- Safety protection of gas circuit
- 230nm grating blazed wavelength, enhanced ultra-violet zone sensitivity
- 1800 g/mm grating density and optimized resolution
Application Fields
- Geology, minerals, metallurgy, steel, non-ferrous metals
- Environmental analysis: air, water quality, soil and solid waste
- Petrochemical industry and light industrial products, crude oil and additional products
- Food, biomedicine and health products
- Building materials (glass, ceramic, paints, etc.)
Technical Specifications | |
---|---|
Optical system: | No lens, concave lens, total reflection, achromatic optical system |
Monochromator: | Czerny-Turner |
Grating: | 1800 lines/mm |
Slit: | auto-switch 0.1,0.2,0.4,0.7, 1.4nm |
Detector: | Photomultiplier |
Light source: | > 8-element lamp turret |
Size/weight: | 130cm x 56cm x 53cm, weight: 165kg |
Graphite Furnace System Specifications | |
Safety protection: | monitoring with automatic shut-off: furnace temperature / graphite tube temperature / cooling flow / system gas pressure |
Graphite Tube: | Pneumatic |
Furnace temperature control: | multiple user modes |
Autosampler | Automatic |
Flame System Specifications | |
Safety protection: | System leakage/ System pressure/ Burner monitoring with automatic shut-off |
Combustion system: | Atomizer, Pure Ti Atomizing Chamber & Burner |
Flame Atomiser positioning | Software controlled horizontal and vertical positioning |
Discharge System: | Automatic |
Additional Details
Address
S-1, 24-C, Sunset Lane No.1, Phase II (ext) DHA, Karachi
Trade & Industrial Type
Industrial Equipment